Recording the nature of the fracture
Upon completion of a pull-off test, the dolly and coated surface should be examined. In addition to pull-off force, many National and International standards such as ASTM D4541 and ISO 4624 require the nature of the fracture to be recorded.
The cohesive and adhesive fracture visual analysis can be recorded using this unit's keypad or touch screen display. Alternatively, the information can be reported using the PosiTector App.
Fractures are visually estimated and recorded as a percent of each. The two layer example below illustrates a cohesive fracture within coating layer 1 and layer 2 (visually estimated at 50% each). Recorded results are included in reports.
This can record fracture information for up to five layers.
Applications
Dolly Size and Stand-Off | 1.97" (50 mm) |
Range | 0.01 MPa (1 psi) |
Accuracy | ±1% full scale |
Cutting Tool | 1.97" (50 mm) hole saw |