ACR 01-0112 SmartReader Plus 1 Data Logger
SKU: ACR 01-0112
Availability: 2 to 3 Weeks ACR ACR 01-0112 Offers Designed for easy monitoring and recording of temperature, this data logger features one internal ambient temperature channel and an input for an optional remote temperature or contact closure sensor. Also, this instrument is ideal for quick and accurate temperature recordings in a variety of applications. Features Two channels Monitor and record temperature Option to monitor remote temperature
Availability: 2 to 3 Weeks
ACR ACR 01-0112 Offers
Designed for easy monitoring and recording of temperature, this data logger features one internal ambient temperature channel and an input for an optional remote temperature or contact closure sensor. Also, this instrument is ideal for quick and accurate temperature recordings in a variety of applications.
Features
- Two channels
- Monitor and record temperature
- Option to monitor remote temperature, resistance or switch status
- Use with a variety of different temperature probes
Applications
- Food process verification
- Pharmaceutical and/or laboratory storage
- Transportation of temperature-sensitive goods
- Equipment run time
- HVAC/R system testing and balancing,
ACR ACR 01-0112 Specifications
Specifications
| Sensor Specifications | |
| Internal Temperature Sensor | NTC Thermistor - 10,000 Ω at 77°F (25°C) |
| Range | -40 to 158˚F (-40 to 70˚C) |
| Accuracy | +/-0.3ºF over the range of 32 to 158ºF (+/-0.2ºC over the range of 0 to 70ºC) |
| Resolution | 0.05ºF (0.03ºC) at 25ºCBetter than 0.12ºF (0.07ºC) between -13º and 158ºF (-25 and 70˚C)Better than 0.23˚F (0.13ºC) between -40 and -13ºF (-40 and -25˚C) |
| Genral Specifications | |
| Number of Channels | Two |
| Sampling Rates | User selectable rates from 25 per second to once every eight hours |
| Memory Size | 32 KB128 KB1.5 MB |
| Software Requirements | TrendReader® Standard 2 |
| PC Requirements | IBM PC or 100% compatible with at least one free serial port, 2 MB of RAM memory, and 2 MB of hard disk space |
| Operating Limits | -40 to 158°F (-40 to 70°C) and 0 to 95% RH (non-condensing) |
| Sampling Methods | Continuous (First-in, First-out)*, Stop When Full (Fill-then-stop) |
| Clock Accuracy | +/-2 seconds per day |
| Mounting | Magnetic backing or locking eyelet |
| Resolution | 12-bit (1 part in 4096) |
| Battery | 3.6 V Lithium |
| Case Material | Noryl® Plastic |
| Size | 4.2 x 2.9 x 0.9" (107 x 74 x 22 mm) |
| Weight | 3.75 oz (110 g) |
What's included with the ACR ACR 01-0112