Guaranteed Fitment
Always the correct part
have questions?
Always the correct part
We know our products
Fast & Reliable
Availability: 5 in Stock
Fluke 393FC OffersWork efficiently with this true-rms clamp meter which features DC power measurement, audio polarity, and visual continuity. Additionally, this meter features wireless transmission and saves test results to a smartphone or the cloud to save time and eliminate data-entry errors.
Features
Applications
Maximum Voltage Between Any Terminal and Earth Ground | 1000 V AC1500 V DC |
Electrical Specifications | Accuracy: Accuracy is specified for 1 year after calibration, at operating temperatures of 64.4 to 82.4°F (18 to 28°C), relative humidity at 0 to 75%. Accuracy specifications take the form of: ±([% of Reading] + [Number of Least Significant Digits])Temperature coefficients: Add 0.1 x specified accuracy for each °C > 64.4°F (28°C) or <82.4°F (18°C) |
AC Current: Jaw | Range: 999.9 AResolution: 0.1 AAccuracy: 2% + 5 digits (10 to 100 Hz); 2.5% + 5 digits (100 to 500 Hz)Crest factor (50/60 Hz):2.5 at 600 A3 at 500 A1.42 at 999.9 AAdd 2% for C.F. >2 |
AC Current: Flexible Current Probe | Range: 999.9 A; 2500 AResolution: 0.1 A (≤999.9 A); 1 A (≤2500 A)Accuracy: 3 % RD + 5 digits (10 to 500 Hz)Crest factor (50/60Hz):2.5 at 1400 A3 at 1100 A1.42 at 2500 AAdd 2% for C.F. >2 |
Position Sensitivity | Distance from optimum:ABCi2500-18 Flex:1.4" (35.6 mm)2" (50.8 mm)2.5" (63.5 mm)Error: ±0.5%±1%±2% |
DC Current | Range: 999.9 AResolution: 0.1 AAccuracy: 2% RD + 5 digits, When using the ZERO function to compensate for offsets |
AC Voltage | Range: 600 V; 1000 VResolution: 0.1 V (≤600 V); 1 V (≤1000 V)Accuracy: 1% RD + 5 digits (20 to 500 Hz) |
DC Voltage | Range: 600 V; 1500Resolution: 0.1 V (≤600 V), 1 V (≤1500 V)Accuracy: 1% RD + 5 digits |
mV DC | Range: 500 mVResolution: 0.1 mVAccuracy: 1% RD + 5 digits |
Amps Frequency: Jaw | Range: 5 to 500 HzResolution: 0.1 HzAccuracy: 0.5% RD + 5 digitsTrigger level:5 to 10 Hz ≥10 A10 to 100 Hz, ≥5 A100 to 500 Hz, ≥10 A |
Amps Frequency: Flexible Current Probe | Range: 5 to 500 HzResolution: 0.1 HzAccuracy: 0.5% RD + 5 digitsTrigger level:5 to 10 Hz ≥10 A10 to 100 Hz, ≥5 A100 to 500 Hz, ≥10 A |
Voltage Frequency | Range: 5 to 500 HzResolution: 0.1 HzAccuracy: 0.5% RD + 5 digitsTrigger level:5 to 20 Hz, ≥5 V20 to 100 Hz, ≥5 V100 to 500 Hz, ≥10 V |
DC Power | Range: 600 kVA (600 V DC range), 1500 kVA (1500 V DC range)Resolution: 0.1 kVA, 1 kVAAccuracy: 2% RD + 2 kVA, 2% RD + 20 kVA |
Resistance | Range: 600 Ω, 6000 Ω, 60 kΩResolution: 0.1 Ω (≤600 Ω), 1 Ω (≤6000 Ω), 0.01 kΩ (≤60 kΩ)Accuracy: 1% RD + 5 digits |
Capacitance | Range: 100 μF, 1000 μFResolution: 0.1 μF (≤100 μF), 1 μF (≤1000 μF)Accuracy: 1% RD + 5 digitsInrush trigger level: 5 A |
Temperature | Operating: 14 to 122°F (-10 to 50°C)Storage: -40 to 140°F (-40 to 60°C) |
Operating Humidity | Non-condensing <50°F (10°C)≤90% RH at 50 to 86°F (10 to 30°C)≤75% RH at 86 to 104°F (30 to 40°C)≤45% RH at 104 to 122°F (40 to 50°C) |
Altitude | Operating: 6561.68' (2000 m)Storage: 39,370.07' (12,000 m) |
Electromagnetic Compatibility (EMC) | International: IEC 61326-1: Portable, Electromagnetic Environment, IEC 61326-2-2 CISPR 11: Group 1, Class A Group 1: Equipment has intentionally generated and/or uses conductively-coupled radio frequency energy that is necessary for the internal function of the equipment itself. Class A: Equipment is suitable for use in all establishments other than domestic and those directly connected to a low-voltage power supply network that supplies buildings used for domestic purposes. There may be potential difficulties in ensuring electromagnetic compatibility in other environments due to conducted and radiated disturbances. Caution: This equipment is not intended for use in residential environments and may not provide adequate protection to radio reception in such environments.Korea (KCC): Class A equipment (Industrial Broadcast and Communications Equipment) Class A: Equipment meets requirements for industrial electromagnetic wave equipment and the seller or user should take notice of it. This equipment is intended for use in business environments and not to be used in homes.USA (FCC): 47 CFR 15 subpart B. This product is considered an exempt device per clause 15.103. |
Safety | General: IEC 61010-1, Pollution Degree 2Measurement:IEC 61010-2-032: CAT III 1500 V / CAT IV 600 VIEC 61010-2-033: CAT III 1500 V / CAT IV 600 V |
Wireless Radio | Radio frequency certification: FCC ID: T68-FBLE, IC: 6627A-FBLEWireless radio frequency range: 2400 to 2483.5 MHzOutput power: <100 mWSIMPLIFIED EU DECLARATION OF CONFORMITY Hereby, Fluke declares that the radio equipment contained in this Product is in compliance with Directive 2014/53/EU. |
Automatic Power Off | 20 minutes |
Display | Dual display with backlight |
Batteries | Two AA IEC LR6 alkaline |
Dimensions | Size: 11.06 x 3.3 x 1.92" (281 x 84 x 49 mm)Jaw opening: 1.33" (34 mm)Flexible current probe diameter: 0.29" (7.5 mm)Flexible current probe cable length (head to electronics connector): 0.07" (1.8 m) |
Weight (With Batteries) | 1.14 lbs (520 g) |
All returns must be authorized. Items returned prior to authorization will not be accepted. For international customers where applicable, customers must ship any returns to us freight and duty prepaid. Return freight charges for internationsl shippments will be the responsibility of the customer.
Defective new equipment may be returned for warrantyrepair/exchange to us within (5) days of the shipment receipt date. After five (5) days, returns will be handled by direct contact and in accordance with the warranty terms of the item's manufacturer. Items that have malfunctioned due to mishandling, electrical overload, etc., will be subject to repair charge.
Non defective new equipment will be authorized for return within five (5) days of the shipment receipt date if the equipment is still in new, re-sellable condition with the original manual(s), accessories, packing material and carton. We reserve the right to reject returnsin which there is no original packaging, missing accessories, damage or visible signs of usage of the item. A restocking fee of up to 25% of purchase price may apply. Special orders and dangerous goods are non-returnable.
(Do not write anytingon the manufacturer's original package or use as a shipping carton. Include all accessories, manuals and warranties.)
Please, include a copy of the packing list or invoice with a reason for the return. Your return authorization number is valid for thirty (30) days.
We can ship to both US and Canada. Note that there are restrictions on some products, and some products cannot be shipped to international destinations.
When you place an order, we will charge a flat rate shipping for your items.
Please also note that the shipping rates for many items we sell are weight-based. The weight of any such item can be found on its detail page. To reflect the policies of the shipping companies we use, all weights will be rounded up to the next full pound.
Not Applicable
Condition: | NA |
Defective new equipment may be returned for warrantyrepair/exchange to us within (5) days of the shipment receipt date. After five (5) days, returns will be handled by direct contact and in accordance with the warranty terms of the item's manufacturer. Items that have malfunctioned due to mishandling, electrical overload, etc., will be subject to repair charge.
Non defective new equipment will be authorized for return within five (5) days of the shipment receipt date if the equipment is still in new, re-sellable condition with the original manual(s), accessories, packing material and carton. We reserve the right to reject returnsin which there is no original packaging, missing accessories, damage or visible signs of usage of the item. A restocking fee of up to 25% of purchase price may apply. Special orders and dangerous goods are non-returnable.
(Do not write anytingon the manufacturer's original package or use as a shipping carton. Include all accessories, manuals and warranties.)
Please, include a copy of the packing list or invoice with a reason for the return. Your return authorization number is valid for thirty (30) days.