GRAPHTEC GL980 OffersFeatures
Simultaneous Sampling - Sampling interval: 1 µs to 1 min (in steps of 1, 2, 5)
GL980 utilizes simultaneous sampling to eliminate slowdown in sampling rate by using multiple A/D converters in simultaneous sampling method. Eight individual A/D converters in each channel sustains the maximum sampling speed for all eightchannels to measure high speed rapid voltage fluctuation and multi-channel vibration measurement.
External Sampling Function - Maximum input frequency: 100 kHz
Sampling of the logger is performed in sync with an external device using an external signal input. *B-513 Input/Output cable for GL is required
Pulse/Logic
Screw Terminal (size M3.5)
Applications
| Main Unit Specifications | |
| Display (LCD) - Size | 7-inch TFT color LCD (WVGA: 800 x 480 dots) |
| Display (LCD) - Information | Waveform in Y-T with digital values, Enlarged waveforms,Digital values and Real-time statistical result values, XY graph |
| Display (LCD) - Language | English, French, German, Spanish, Russian, Chinese, Korean, Japanese |
| Interface to PC - Type | Ethernet (10 BASE-T/100 BASE-TX), USB2.0 |
| Interface to PC - Function | Data transfer to PC (up to 1 ms sampling), Control command to GL2000 |
| Interface to PC - Ethernet functions | Web server function, FTP server function, NTP client function, DHCP client function, Email send function |
| Interface to PC - USB function | USB mode (File transfer and deletion from built-in flash and SD on GL2000) |
| Trigger function - Trigger action | Start or stop capturing data by triggering |
| Trigger function - Trigger source (Start) | Off, Measured signal, Alarm, External, Scheduled time, Scheduled day, Elapsed time |
| Trigger function - Trigger source (Stop) | Off, Measured signal, Alarm, External, Scheduled time, Scheduled day, Elapsed time |
| Trigger function - Combination | Level OR, Level AND, Edge OR, Edge AND |
| Trigger function - Threshold (Analog) | High or Low in level mode, Rising or Falling in edge mode, Window-in, Window-out |
| Trigger function - Threshold (Logic) | H or L (4-channel signal pattern) |
| Trigger function - Threshold (Pulse) | High or Rising, Low or Falling, Window-in, Window-out |
| Trigger function - Repeat action | Off, On (Re-armed automatically) |
| Trigger function - Trigger hold out | Hold off repeat action in specified period |
| Trigger function - Mode | Previous start to next start, previous stop to next start |
| Trigger function - Defection accuracy | ± 0.5 % of measurement range |
| Trigger function - Pre-trigger | Up to the number of capturing data points (max. 4000000)specified in built-in RAM (only when built-in RAM is used) |
| Alarm function - Alarm action | Displays and outputs a signal when alarm is detected |
| Alarm function - Threshold (Analog input) | High, Low, Window-in, Window-out |
| Alarm function - Threshold (Logic input) | H or L (signal in each channel) |
| Alarm function - Threshold (Pulse input) | High or Rising, Low or Falling, Window-in, Window-out |
| Alarm function - Combination | OR (Source channel can be assigned with OR condition to output port) |
| Alarm function - Detection cycle | Link with analog sampling |
| Alarm function - Alarm holding | On or Off |
| Alarm function - Detection accuracy | ± 0.5 % of measurement range |
| Storage device - Built-in RAM | Four million samples for each channel |
| Power consumption (AC adapter in 240 V AC) | Approx. 48 V A (66 V A while charging battery) with disabling screen saver |
| Power consumption (DC drive - 24 V) | Approx. 0.6 A (0.9 A while charging battery) with disable screen saver |
| Power consumption (DC drive - 12 V) | Approx. 1.22 A (Cannot charge battery) with disable screen saver |
| Power consumption (DC drive - 8.5 V) | Approx. 1.81 A (Cannot charge battery) with disable screen saver |
| External dimensions [WxHxD] | Approx. 260 x 161 x 83 mm (with the cover) |
| Weight | Approx. 1.7 kg (the cover is attached, AC adapter and battery are not included) |
| Vibration resistance | Compatible with JIS Vibration test method for automobile Type 1 Class A(Vibration durability test: 5 m/s²) |
| Analog Unit Specifications | |
| Number of input channels | 8 channels |
| Type of input terminal | Isolated BNC connector and Screw terminal (M3.5 screw) (*6) |
| Input method | All channels isolated unbalanced input, Simultaneous sampling |
| Sampling speed (interval) | 1 M Samples/s to 1 Sample/min (1 µs to 1 min) and External (*7) |
| Sampling interval | 1, 2, 5, 10, 20, 50, 100, 200, 500 µs |
| Frequency response | DC to 200 kHz (within +1/-4 dB) |
| Measurement range - Voltage (DC) | 20, 50, 100, 200, 500 mV, 1, 2, 5, 10, 20, 50, 100, 200, 500, 1000 V, and 1-5V F.S. * Max. rated safety voltage: ±600 V DC |
| Measurement range - DC-RMS (DC coupling and rms value meas.) | 10, 25, 50, 100, 250, 500 mV rms, 1, 2, 5, 10, 25, 50, 100, 250, 500, 1000 V rms F.S. * Max. rated safety voltage: 600 V rms |
| Measurement range - Temperature | Thermocouple: K, J, E, T, R, S, B, N, W (WRe5-26) |
| Measurement range - Humidity | 0 to 100 % RH - using the humidity sensor (option B-530) |
| Filter (Low pass) | Off, Line (1.5 Hz), 5, 50, 500 Hz, 50 kHz (at -3dB, -6dB/oct) |
| A/D converter | 16-bit (effective resolution: 1/40000 of the measuring full range) |
| Measurement accuracy - Voltage (DC) | ± 0.25% of Full Scale |
| Measurement accuracy - Voltage (RMS) | ± 1.5% of Full Scale (Sine wave in 20 Hz - 10 kHz) |
| Measurement accuracy - Temperature (Thermocouple) | Type: R/S, B, K, E, T, J, N, W |
| R.J. Compensation | Internal or External |
| Burnout | Detecting burnout of Thermocouple with menu operation in free-run mode |
| Input impedance | 1 M&Ohm; ±5% |
| Signal source impedance | up to 1 k&Ohm; |
| Maximum input voltage | Between(+) - (-) terminals: 20 mV to 2 V range: 30 V DC, 5 V to 500 V range: 500 V DCBetween channels (-) - (-) terminals: 60 V P-PBetween channel - GND: 60 V P-P |
| Isolation Resistance | Min. 50 M&Ohm; (at 500 V DC) with between input and GND |
| Common-mode rejection ratio | Min. 90 dB (50/60 Hz, signal source impedance: max. 300 &Ohm;) |
| Signal-Noise Ratio (S/N) | 20 mV range: -40 dB (when input terminals + and - are shorted)Other range: -50 dB (when input terminals + and - are shorted) |
All returns must be authorized. Items returned prior to authorization will not be accepted. For international customers where applicable, customers must ship any returns to us freight and duty prepaid. Return freight charges for internationsl shippments will be the responsibility of the customer.
Defective new equipment may be returned for warrantyrepair/exchange to us within (5) days of the shipment receipt date. After five (5) days, returns will be handled by direct contact and in accordance with the warranty terms of the item's manufacturer. Items that have malfunctioned due to mishandling, electrical overload, etc., will be subject to repair charge.
Non defective new equipment will be authorized for return within five (5) days of the shipment receipt date if the equipment is still in new, re-sellable condition with the original manual(s), accessories, packing material and carton. We reserve the right to reject returnsin which there is no original packaging, missing accessories, damage or visible signs of usage of the item. A restocking fee of up to 25% of purchase price may apply. Special orders and dangerous goods are non-returnable.
(Do not write anytingon the manufacturer's original package or use as a shipping carton. Include all accessories, manuals and warranties.)
Please, include a copy of the packing list or invoice with a reason for the return. Your return authorization number is valid for thirty (30) days.
We can ship to both US and Canada. Note that there are restrictions on some products, and some products cannot be shipped to international destinations.
When you place an order, we will charge a flat rate shipping for your items.
Please also note that the shipping rates for many items we sell are weight-based. The weight of any such item can be found on its detail page. To reflect the policies of the shipping companies we use, all weights will be rounded up to the next full pound.
Not Applicable
| Condition: | NA |
Defective new equipment may be returned for warrantyrepair/exchange to us within (5) days of the shipment receipt date. After five (5) days, returns will be handled by direct contact and in accordance with the warranty terms of the item's manufacturer. Items that have malfunctioned due to mishandling, electrical overload, etc., will be subject to repair charge.
Non defective new equipment will be authorized for return within five (5) days of the shipment receipt date if the equipment is still in new, re-sellable condition with the original manual(s), accessories, packing material and carton. We reserve the right to reject returnsin which there is no original packaging, missing accessories, damage or visible signs of usage of the item. A restocking fee of up to 25% of purchase price may apply. Special orders and dangerous goods are non-returnable.
(Do not write anytingon the manufacturer's original package or use as a shipping carton. Include all accessories, manuals and warranties.)
Please, include a copy of the packing list or invoice with a reason for the return. Your return authorization number is valid for thirty (30) days.